Pengukuran MTF (Modulation Transfer Function) berdasarkan LSF (Line Spread Function) dan PSF (Point Spread Function) pada pesawat PET/CT dan SPECT/CT

Sharah Nataz Shilfa, M. Roslan Abdul Gani, Intan Apriliani Syaridatul Mu’minah, Fery Ardiansyah, Lukmanda Evan Lubis, Djarwani Soeharso Soejoko

Abstract


This study aims to know the value of MTF obtained by using two methods: LSF (Line Spread Function) and PSF (Point Spread Function) on PET/CT and SPECT/CT. Measurement done with LSF method performed by thin flood source 18F or 99mTc 12 mCi, while PSF using thin line source 18F atau 99mTc 2 mCi with diameter 0.5 mm. The MTF calculations are performed based on image profile, for PET/CT assisted by 3 filter types and 4 reconstruction algorithms while SPECT/CT based on collimator variation. MTF calculation method have done based on Tsujioka et al. (2014) by combining the ImageJ, MATLAB, and Origin spreadsheet. System performance is determined at 10% MTF. The measurement results on PET/CT showed MTF highest value obtained by the Butterworth filter and reconstruction algorithm Iterative, Iterative 3D, and Backprojection with identical MTF value (0.155 cycles/mm) for LSF method, while PSF method by using Butterworth filter and Backprojection reconstruction algorithm obtained value 0.15 cycles/mm. In SPECT/CT, the second detector with LEHR collimator give MTF value to the LSF method and PSF method sequently 0.104 cycles/mm and 0075 cycles/mm. Based on the advantages and disadvantages of both methods, generally the PSF method is recommended as a spatial resolution QC program for nuclear medicine modalities.


Full Text:

PDF

Refbacks

  • There are currently no refbacks.


©2017 (onwards) Aliansi Fisikawan Medik Indonesia / Indonesian Association of Physicists in Medicine

Print ISSN: 2355-2727 | Online ISSN: 2355-2719